Retrograde nailing versus locking plate osteosynthesis of proximal humeral fractures: a biomechanical study
2011
Background In locking plate osteosynthesis of proximal humeral fractures, secondary varus malalignment is a specific complication. Retron nails (Tantum AG, Neumunster, Germany), among others, have been designed to improve medial support of the calcar humeri. The aim of our biomechanical study was to examine whether Retron nails provide increased stiffness for axial loads and adequate stiffness for torsional loads when compared with Philos plates (Synthes AG, Umkirch, Germany). Materials and methods Twenty-two fresh-frozen paired humeri were collected. After potting the specimens, intact bones were exposed to sinusoidal axial (10-120 N) and torsional (±2.5 Nm) loading for 8 cycles to calcula…
Influence of the Screw Positioning on the Stability of Locking Plate for Proximal Tibial Fractures: A Numerical Approach
2020
Tibial fractures are common injuries in people. The proper treatment of these fractures is important in order to recover complete mobility. The aim of this work was to investigate if screw positioning in plates for proximal tibial fractures can affect the stability of the system, and if it can consequently influence the patient healing time. In fact, a more stable construct could allow the reduction of the non-weight-bearing period and consequently speed up the healing process. For that purpose, virtual models of fractured bone/plate assemblies were created, and numerical simulations were performed to evaluate the reaction forces and the maximum value of the contact pressure at the screw/bo…